Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode either manually or remove the crucible and press the analyze button to clean with an automatic cleaner if applicable.
Add approximately 0.05 g of *** Graphite Powder to a 782-720 Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Software Features & Benefits
Cornerstone brand software is divided into four main sections—Analysis, Diagnostics, Settings, and Instrument—for simplified navigation and organization. Toolbars, sliders, and drop-down menus makeit easy to set parameters for calibration and data processing. The software also includes real-time ambient monitors, with fully animated system diagrams. Advanced diagnostic features include a thorough digital on-board manual, maintenance animations, photo illustrations, and screen captures that quickly provide the direction needed without having to refer to multiple manuals. also Cornerstone supports extended data archiving and flexible reporting capabilities. 新疆氧氮?dú)浞治鰞x采購氧氮?dú)浞治鰞x怎么選?歡迎聯(lián)系上海禹重科技,我司產(chǎn)品質(zhì)量保證,歡迎咨詢!
Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Instrument Highlights and Features
Optional Features
Optional boom-mounted touch-screen interface package (M) promotes an ergonomic workspace
Optional Performance Package (P) adds a three-stage carrier gas purification system, and a dual-loop gas dose module
Optional Dual Cooling Package (D) adds an integratedliquid-to-air radiator with dual cooling fans. This DC
option gives the user the choice of liquid-to-liquid cooling, liquid-to-air cooling, or a combination of both
Optional integrated autocleaner (H) automatically cleans the upper electrode and lower electrode tip
Optional 20-sample shuttle loader available
配置SmartLine? Remote遠(yuǎn)程診斷功能。
Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of three Standard reps.
Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.
簡化日常分析所需的爐頭進(jìn)樣部分機(jī)械維護(hù)工作。安徽LECO分析儀廠家直銷
可選六軸可旋轉(zhuǎn)觸摸屏幕,增強(qiáng)了操作的工效和直觀性。廣東美國分析儀廠家報(bào)價(jià)
Procedure - Solid Samples
Press the Analyze button on the instrument screen again, the loading head slide-block will close and the lower electrode will open.
Clean the upper and lower electrode manually, or, if applicable, remove the crucible and press the analyze button to clean with the automatic cleaner
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically